Some hearing instruments are equipped with technology that allows the instrument to re-position highfrequency
input information in a lower frequency output register. Using Audioscan equipment running V3.4 software
or higher, this functionality can be tested using one of three modied versions of the Speech-std (1) test
stimulus (Speech4000, Speech5000 and Speech6300) which are available in the Speechmap stimulus menu.
In each of these modied test stimuli, the 1/3 octave band
levels above 1000Hz are reduced by 30dB, except for an
isolated 1/3 octave band centered at the frequency indicated
in the selected test stimulus’ name. With these
reduced band levels, the resulting input LTASS produces a
distinct “cavity” between 1000Hz and the selected high
frequency band, as seen at left.